Microlife Microlife
  • Voor de consument
    • BP A7 Touch BT_half_cuff Bloeddrukmeters
    • NC200_front Thermometers
    • BreastPump_BC200-front Baby Care
    • NEB NANObasic_half-acc Ademhalingszorg
    • FH_422 Warmtetherapie
    • WS 200 BT_front Gewicht
    • Microlife WS 80-N half Weight
    • Software_downloads_allg Software
  • Voor de professional
    • WatchBP Office ABI_front_cuff WatchBP Office
    • WatchBP O3 Ambulatory AFIB_front_cuff WatchBP O3
    • WatchBP Home A_front_cuff WatchBP Home
    • Cuff-Ankle Manchetten en onderdelen
    • Software Software
    • PP-In-Use WBP Office_Inter-Arm-Difference Proefplaatsing
  • Klinische Validaties
  • Technologieën
  • Health Magazine
    • icon_heart_blue Bloeddruk
    • icon_Temperature Koorts
    • icon_airway-treatment Luchtwegaandoening
    • icon_even_heat_blue Warmtetherapie
  • Support
    • BP A2 Basic startseite Hulp bij producten
    • Software_downloads_allg Software voor thuisgebruik
    • Software_downloads_WBP Software voor de professional
  • Over Microlife
    • aboutus Over ons
    • Events Nieuws & Events
    • contact Contact
    • History Geschiedenis
  • Developers
NEDERLAND
Nederlands
Europe
  • EUROPE - English
  • CZECHIA - Čeština
  • DANMARK - Dansk
  • DEUTSCHLAND - Deutsch
  • ESPAÑA - Español
  • FRANCE - Français
  • HRVATSKA - Hrvatski
  • ITALIA - Italiano
  • LATVIJA - Latviešu
  • LIETUVA - Lietuvių
  • MAGYARORSZÁG - Magyar
  • NEDERLAND - Nederlands
  • ÖSTERREICH - Deutsch
  • POLSKA - Polski
  • PORTUGAL - Português
  • SCHWEIZ - Deutsch
  • SLOVENSKO - Slovenčina
  • SLOVENIJA - Slovenščina
  • SUOMI - Suomi
  • SCHWEIZ - Français
  • UK - English
  • ΕΛΛΆΔΑ - Ελληνικά
  • БЪЛГАРИЯ - Български
  • СРБИЈА - Српски
  • УКРАЇНА - Українська
  • TÜRKIYE - Türkçe
  • РОССИЯ - Русский
  • БЕЛАРУСЬ - Русский
Asia
  • ASIA PACIFIC - English
  • VIỆT NAM - Tiếng Việt
  • ประเทศไทย - ไทย
  • 中国 - 中文
  • 台灣 - 中文
Middle East and North Africa
  • MENA - Arabic عربى
USA, Latin America
  • US - English
  • AMÉRICA DEL SUR - Español
  • COLOMBIA - Español
  • MÉXICO - Español
Menu
Header Microlife & WatchBP Clinical Studies Flow

Triage tests for identifying atrial fibrillation in primary care: a diagnostic accuracy study comparing single-lead ECG and modified BP monitors

Home > Voor de professional > Validaties & Klinische studies > Triage tests for identifying atrial fibrillation in primary care: a diagnostic accuracy study comparing single-lead ECG and modified BP monitors, Kearley, 2014
Share

2014, Karen Kearley, Mary Selwood, Ann Van den Bruel, Matthew Thompson, David Mant, FD Richard Hobbs, David Fitzmaurice, Carl Heneghan

Abstract

Objective:
New electronic devices offer an opportunity within routine primary care settings for improving the detection of atrial fibrillation (AF), which is a common cardiac arrhythmia and a modifiable risk factor for stroke. We aimed to assess the performance of a modified blood pressure (BP) monitor and two singlelead ECG devices, as diagnostic triage tests for the detection of AF.

Setting:
6 General Practices in the UK.

Participants:
1000 ambulatory patients aged 75 years and over.

Primary and secondary outcome measures:
Comparative diagnostic accuracy of modified BP monitor and single-lead ECG devices, compared to reference standard of 12-lead ECG, independently interpreted by cardiologists. 

Results:
A total of 79 participants (7.9%) had AF diagnosed by 12-lead ECG. All three devices had a high sensitivity (93.9–98.7%) and are useful for ruling out AF. WatchBP is a better triage test than Omron autoanalysis because it is more specific—89.7% (95% CI 87.5% to 91.6%) compared to 78.3% (95% CI 73.0% to 82.9%), respectively. This would translate into a lower follow-on ECG rate of 17% to rule in/rule out AF compared to 29.7% with the Omron text message in the study population. The overall specificity of single-lead ECGs analysed by a cardiologist was 94.6% for Omron and 90.1% for Merlin.

Conclusions:
WatchBP performs better as a triage test for identifying AF in primary care than the singlelead ECG monitors as it does not require expertise for interpretation and its diagnostic performance is comparable to single-lead ECG analysis by cardiologists. It could be used opportunistically to screen elderly patients for undiagnosed AF at regular intervals and/or during BP measurement.

Download Abstract as PDF-file